#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2211D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.12.2023, Control date: 12.12.2023,
Control # 4
No serum blank (average) : 0.139
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.008 |
0.007 |
-0.005 |
0.006 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.073 |
0.114 |
0.084 |
0.112 |
0.096 |
4 |
Weak positive serum
(N° 9650-07) |
0.369 |
0.377 |
0.400 |
0.396 |
0.386 |
18 |
Positive control
serum
(N° 9650-08) |
2.106 |
2.232 |
2.163 |
2.246 |
2.187 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.187 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.139 |
Yes |
Comments:
Good for use.