#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2211D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.12.2023, Control date: 05.04.2023,
Control # 3
No serum blank (average) : 0.121
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.003 |
-0.003 |
-0.001 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.073 |
0.085 |
0.079 |
0.078 |
0.079 |
4 |
Weak positive serum
(N° 9650-07) |
0.344 |
0.331 |
0.336 |
0.355 |
0.342 |
16 |
Positive control
serum
(N° 9650-08) |
2.148 |
2.151 |
2.109 |
2.223 |
2.158 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.158 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.121 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|