#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2129D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.07.2023, Control date: 24.11.2021,
Control # 1
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.005 |
-0.005 |
0.005 |
-0.005 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.102 |
0.090 |
0.105 |
0.086 |
0.096 |
4 |
Weak positive serum
(N° 9650-07) |
0.383 |
0.348 |
0.355 |
0.325 |
0.353 |
14 |
Positive control
serum
(N° 9650-08) |
2.505 |
2.557 |
2.447 |
2.453 |
2.491 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.491 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.