#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2128T
Results of quality control
Kit: Bordier Affinity Products SA, Toxocara canis IgG ELISA, Article 9200
Expiry date: 07.07.2023, Control date: 06.05.2022,
Control # 2
No serum blank (average) : 0.121
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.001 |
0.002 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.002 |
0.004 |
0.007 |
0.023 |
0.009 |
0 |
Weak positive serum
(N° 9200-07) |
0.972 |
0.983 |
1.072 |
1.122 |
1.038 |
53 |
Positive control serum
(N° 9200-08) |
1.782 |
1.897 |
1.968 |
2.166 |
1.954 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
1.954 |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 8 % |
0 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.121 |
Yes |
Comments:
Good for use.