#9310
Echinococcus
multilocularis (Em18) IgG ELISA
ELISA kit for alveolar echinococcosis patient follow-up in humans
Lot 2109F
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus multilocularis (Em18)
IgG ELISA, Article 9310
Expiry date: 10.11.2022, Control date: 17.03.2021,
Control # 1
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.009 |
-0.004 |
0.011 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9310-06) |
-0.001 |
0.012 |
0.005 |
0.013 |
0.007 |
0 |
Weak positive serum
(N° 9310-07) |
0.126 |
0.122 |
0.161 |
0.160 |
0.142 |
6 |
Positive control serum
(N° 9310-08) |
2.096 |
2.231 |
2.249 |
2.512 |
2.272 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9310-08 |
> 1.200 |
2.272 |
Yes |
Absorbance of negative serum 9310-06, as % of positive serum |
< 5 % |
0 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|