#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2107D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica IgG ELISA, Article
9650 .
Expiry date: 10.11.2022, Control date: 03.09.2021,
Control # 2
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.005 |
0.003 |
-0.006 |
0.008 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.103 |
0.131 |
0.077 |
0.106 |
0.104 |
5 |
Weak positive serum
(N° 9650-07) |
0.294 |
0.288 |
0.258 |
0.268 |
0.277 |
13 |
Positive control
serum
(N° 9650-08) |
2.093 |
2.368 |
2.006 |
2.092 |
2.140 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.140 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.