#9650
Fasciola
hepatica (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2019D
Results of quality control
Kit: Bordier Affinity Products SA, Fasciola hepatica ELISA, Article
9650 .
Expiry date: 30.04.2022, Control date: 01.03.2020,
Control # 2
No serum blank (average) : 0.125
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.004 |
0.003 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.059 |
0.056 |
0.045 |
0.065 |
0.057 |
3 |
Weak positive serum
(N° 9650-07) |
0.224 |
0.252 |
0.238 |
0.282 |
0.249 |
12 |
Positive control
serum
(N° 9650-08) |
2.079 |
2.287 |
2.063 |
2.159 |
2.147 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.147 |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 9 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.125 |
Yes |
Comments:
Good for use.