#9400
Acanthocheilonema
viteae (Onchocercidae)
ELISA kit for the diagnosis of human filariasis
Lot 2017V
Results of quality control
Kit: Bordier Affinity Products SA, Acanthocheilonema viteae ELISA, Article 9400 .
Expiry date: 21.02.2022, Control date: 29.06.2020,
Control # 1
No serum blank (average) : 0.129
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.004 |
-0.001 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9400-06) |
0.108 |
0.115 |
0.108 |
0.129 |
0.115 |
5 |
Weak positive serum
(N° 9400-07) |
1.134 |
1.103 |
1.120 |
1.065 |
1.105 |
44 |
Positive control
serum
(N° 9400-08) |
2.460 |
2.527 |
2.436 |
2.555 |
2.494 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9400-08 |
> 1.200 |
2.494 |
Yes |
Absorbance of negative serum 9400-06, as % of positive serum |
< 12 % |
5 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.129 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|