#9350
Echinococcus
granulosus
ELISA kit for the diagnosis of echinococcoses in humans
Lot 2014G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus ELISA, Article 9350
Expiry date: 03.02.2022, Control date: 01.12.2020,
Control # 2
No serum blank (average) : 0.163
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.016 |
0.014 |
-0.015 |
0.017 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.018 |
0.081 |
0.007 |
0.069 |
0.043 |
2 |
Weak positive serum
(N° 9350-07) |
0.950 |
0.972 |
0.880 |
0.972 |
0.943 |
49 |
Positive control serum
(N° 9350-08) |
1.830 |
2.117 |
1.812 |
1.956 |
1.928 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
1.928 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.163 |
Yes |
Comments:
Good for use.