#9350
Echinococcus
granulosus
ELISA kit for the diagnosis of echinococcoses in humans
Lot 2014G
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus granulosus ELISA, Article 9350
Expiry date: 03.02.2022, Control date: 22.06.2020,
Control # 1
No serum blank (average) : 0.131
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
-0.001 |
0.008 |
-0.005 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.041 |
0.054 |
0.062 |
0.066 |
0.056 |
3 |
Weak positive serum
(N° 9350-07) |
1.146 |
1.118 |
1.158 |
1.145 |
1.142 |
53 |
Positive control serum
(N° 9350-08) |
2.052 |
2.172 |
2.102 |
2.332 |
2.165 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.165 |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 12 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.131 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|