#9300
Echinococcus
multilocularis (Em2-Em18)
ELISA kit for the diagnosis of alveolar echinococcosis in humans
Lot 2012M
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus multilocularis
Em2-Em18 ELISA, Article
9300
Expiry date: 06.01.2022, Control date: 10.01.2022,
Control # 4
No serum blank (average) : 0.128
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.010 |
0.000 |
0.009 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.009 |
-0.038 |
0.018 |
-0.005 |
-0.004 |
0 |
Weak positive serum
(N° 9300-07) |
0.548 |
0.603 |
0.578 |
0.629 |
0589 |
25 |
Positive control serum
(N° 9300-08) |
2.180 |
2.419 |
2.245 |
2.500 |
2.336 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9300-08 |
> 1.200 |
2.336 |
Yes |
Absorbance of negative serum 9300-06, as % of positive serum |
< 12 % |
0 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.128 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|