#9300
Echinococcus
multilocularis (Em2-Em18)
ELISA kit for the diagnosis of alveolar echinococcosis in humans
Lot 2012M
Results of quality control
Kit: Bordier Affinity Products SA, Echinococcus multilocularis
Em2-Em18 ELISA, Article
9300
Expiry date: 06.01.2022, Control date: 03.05.2021,
Control # 3
No serum blank (average) : 0.133
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.003 |
-0.002 |
0.001 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.033 |
0.027 |
0.028 |
0.031 |
0.030 |
1 |
Weak positive serum
(N° 9300-07) |
0.618 |
0.601 |
0.627 |
0.665 |
0628 |
28 |
Positive control serum
(N° 9300-08) |
2.168 |
2.308 |
2.210 |
2.319 |
2.251 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9300-08 |
> 1.200 |
2.251 |
Yes |
Absorbance of negative serum 9300-06, as % of positive serum |
< 12 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.133 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|