#9750
Trichinella spiralis
ELISA kit for the diagnosis of trichinellosis in humans
Lot 2008R
Results of quality control
Kit: Bordier Affinity Products SA, Trichinella spiralis ELISA, Article
9750 .
Expiry date: 11.11.2021, Control date: 15.11.2021,
Control # 4
No serum blank (average) : 0.123
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
0.000 |
0.005 |
-0.005 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.033 |
0.043 |
0.039 |
0.045 |
0.040 |
2 |
Weak positive serum
(N° 9750-07) |
0.199 |
0.246 |
0.216 |
0.234 |
0.224 |
10 |
Positive control serum
(N° 9750-08) |
2.280 |
2.309 |
2.188 |
2.470 |
2.312 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.312 |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 10 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.123 |
Yes |
Comments:
Good for use.